Datamax H-4606X Manuale dell'Operatore Pagina 95

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 94
H-Class 85
Advanced Entry Calibration (continued)
Step
Action Displayed Message Comment
L
Use the buttons to set the
Gain Number determined
in the previous step. Press
ENTER to enable the
setting.
TRAN SENSOR GAIN
yyy (0 - 31)
18
This example uses Gain
Number 18.
M
Complete a table (see
example below) using new
measurements, as
follows:
(A) Raise the Printhead
Assembly. Place the
label over the Media
Sensor then lower and
latch the Printhead
Assembly. Record the
sensor reading as P.
(B) Raise the Printhead
Assembly. Place the
liner, notch, or mark
over the Media Sensor
then lower and latch
the Printhead
Assembly. Record the
sensor reading as G
(or M).
(C) Raise the Printhead
Assembly. Remove all
media from the Media
Sensor then lower and
latch the Printhead
Assembly. Record the
sensor reading as E.
TRAN SENSOR GAIN
yyy (0 - 31)
18
Where ‘yyy’ is a
numerical value
representing the current
sensor reading.
;
The re-sampled
values may differ
from those
previously noted.
This is normal; do
not readjust the
Media Sensor.
Selected Gain Table
Gain Number Paper Gap (or Mark) Empty
18 173 42 9
Vedere la pagina 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 185 186

Commenti su questo manuale

Nessun commento