82 H-Class
Advanced Entry Calibration (continued)
Step
Action Displayed Message Comment
F
Use the buttons to set the
Gain Number to 00 and
then press ENTER.
Record the sensor reading
as a Label Value for Gain
Number 00 in a table (32
rows by four columns,
with headings similar to
those shown below.)
TRAN SENSOR GAIN
yyy (0 - 31)
00
This is the Label Value for
a gain setting of 00,
where ‘yyy’ represents
the current sensor
reading.
Sample Calibration Table
Gain Number Label Value TOF Value Difference Value
00 252
01
02
31
Step
Action Displayed Message Comment
G
Use the buttons to
increment the Gain
Number by one and then
press ENTER. Record the
Label Value. Repeat this
process for each Gain
Number.
TRAN SENSOR GAIN
yyy (0 - 31)
01
These are Label Values,
where ‘yyy’ represents
the current sensor
reading.
Sample Calibration Table
Gain Number Label Value TOF Value Difference Value
00 252
01 250
02 248
… …
31 09